12 lead ecg
From 2nd Book
Jump to navigationJump to search
12 lead ecg [∞]
Since each ECG lead is derived from a subset of electrodes, we can reposition the three electrodes in a systematic sequence to collect all necessary data for a full diagnostic ECG. Here’s how we can do it step by step:
Step 1: Record the Standard Limb Leads (I, II, III)
- Placement:
- Right Hand (RA)
- Left Hand (LA)
- Left Leg (LL)
- Leads Recorded:
- Lead I (LA - RA)
- Lead II (LL - RA)
- Lead III (LL - LA)
Step 2: Derive the Augmented Leads (aVR, aVL, aVF)
- These leads are mathematically derived from the standard limb leads and do not require repositioning.
Step 3: Record the Chest Leads (V1–V6)
Since we only have three electrodes available at any given time, we need to move one of the limb electrodes onto the chest while maintaining a reference point.
- V1 Recording
- Right Hand (RA) stays on the right arm.
- Left Hand (LA) is moved to V1 position (4th intercostal space, right sternal border).
- Left Leg (LL) remains as a reference (ground).
- V2 Recording
- Move the LA electrode from V1 to V2 (4th intercostal space, left sternal border).
- V3 Recording
- Move LA to V3 (midway between V2 and V4).
- V4 Recording
- Move LA to V4 (5th intercostal space, midclavicular line).
- V5 Recording
- Move LA to V5 (same level as V4, anterior axillary line).
- V6 Recording
- Move LA to V6 (same level as V4 and V5, midaxillary line).
Final Considerations
- Each recording step captures one or more leads at a time.
- The RA and LL electrodes remain relatively fixed to maintain proper signal integrity while the LA electrode is repositioned systematically across the chest.
- Once all signals are collected, the full 12-lead ECG can be reconstructed.
This method allows a 3-wire ECG system to capture all 12 leads by repositioning the electrodes in a series of recordings.